Glossary of terms
This glossary includes some of the terms and abbreviations used regularly in the pharmaceutical water and sterilisation arenas. If there are other terms you feel would help please let us know and we will include them in this list.
| Active
drain: |
A drain, which is situated at the lowest part of the sterilizer chamber to control the discharge of air/non-condensable gases or air and condensate from the sterilizer chamber. |
| Air
detector: |
A device used to determine that sufficient air or other non-condensable gases have been removed from the chamber. |
| Air
removal: |
Removal of air from the sterilizer chamber and sterilizer load sufficient to facilitate steam penetration. |
| Automatic
controller: |
Device that, in response to pre-determined cycle variables, operates the sterilizer sequentially through the required stages of the cycle(s). |
| Automatic
control test: |
A test designed to show that the operating cycle functions correctly as evidenced by the values of the cycle variables indicated and recorded by the instruments fitted to the sterilizer. |
| Bioburden: |
Population of viable micro-organisms on a product and/or package. |
| Biological
indicator: |
An inoculated carrier contained within its primary pack ready for use. |
| Calibration: |
The set of operations that establish, under specified conditions, the relationship between values of a quantity indicated by a measuring instrument or measuring system, or values represented by a material measure or a reference material, and the corresponding values realised by standards. |
| Chamber
depth: |
Depth
of the sterilizer chamber which is available for the sterilizer load. |
| Chamber
height: |
Height
of the sterilizer chamber which is available for the sterilizer load. |
| Chamber
width: |
Width
of the sterilizer chamber which is available for the sterilizer load |
| Chamber
temperature: |
Lowest
temperature prevailing in the sterilizer chamber. |
| Chemical
Indicator: |
A
device designed to show, usually by a change of colour; whether specified
values of one or more cycle variables nave been attained. |
| Disinfection: |
A
process used to reduce the number of viable micro-organisms in a load
but which may not necessarily inactivate some viruses and bacterial
spores. |
| Dedicated
steam supply: |
Supply
of steam produced for a sterilizer, or group of sterilizers, by a dedicated
generator. |
| Door: |
Lid
or similar device provided as a means of closing and sealing the sterilizer
chamber. |
| Double
ended sterilizer: |
Sterilizer
in which there is a door at each end of the sterilizer chamber. |
| Dryness
value: |
A
dimensionless quantity approximating to the dryness fraction, derived
to determine whether steam is of the correct dryness for sterilization
purposes. A dryness value of 1.0 represents dry saturated steam. |
| Dry
saturated steam: |
Steam
with a temperature and pressure corresponding to the vaporisation curve
of water.
NOTE:
This is an ideal condition, which can deviate towards either superheated
steam or to wet steam. This deviation is quantified by the determination
of the Dryness Value. |
| D-value: |
Decimal
reduction value (for biological indicators". The time in minutes
required to secure inactivation of 90% of the test organisms under stated
exposure conditions. |
| Equilibration
time: |
Period
which elapses between the attainment of the sterilization temperature
in the sterilizer chamber and the attainment of the sterilization temperature
at all points within the load. |
| Fail
safe: |
Attribute
of sterilizer design, component or its associated services that minimises
a possible safety hazard. |
| Fault: |
Recognition
by the automatic controller that the pre-set cycle variables for the
sterilization cycle have not been attained. |
| Holding
time: |
Period
for which the temperature of all points within the sterilizer is held
within the sterilization temperature band.
NOTE:
The holding time follows immediately after the equilibration time.
The extent of the holding time is related to the sterilization temperature |
| Inoculated
carrier: |
A
carrier on which a defined number of test organisms has been deposited. |
| Installation
test: |
Series
of checks and tests performed after installation of the sterilizer in
the place of use. |
| Loading
door: |
Door
in a double ended sterilizer through which the sterilizer load is put
into the sterilizer chamber prior to sterilization. |
| Medical
device: |
The
definition given in EN 46001 applies. |
| Moist
heat: |
Heat
that is derived from water, either as a liquid or as steam under pressure. |
| Parametric
release: |
Declaring
product as 'sterile' based on physical process data rather than on the
basis of sample testing or biological indicator results. |
| Performance
qualification: |
Obtaining
and documenting evidence that the equipment as commissioned will produce
acceptable product when operated in accordance with the process specification. |
| Plateau
period: |
Equilibration
time plus the holding time. |
| Primary
standard: |
Standard
that is designated or widely acknowledged as having the highest metrological
qualities and whose values are accepted without reference to other standards. |
| Product
compatibility: |
Ability
of the sterilization cycle to achieve the intended results without detrimental
effect on the product. |
| Non-condensable
gas: |
Air
and other gas which will not condense under the conditions of steam
sterilization. |
| Pressure
vessel: |
A
vessel describing the sterilizer chamber, jacket (if fitted), door(s)
and components that are in permanent connection with the sterilizer
chamber. |
| Recommissioning: |
Procedure
to confirm that the sterilizer functions in accordance with its specification
and that data established during commissioning remain valid. |
| Reference
measurement point: |
Reference
point for which documented evidence is available to demonstrate that
it has a known relationship to the temperature of the coolest part of
the sterilizer chamber. |
| Reference
load: |
Specified
load made up to represent the most difficult combination of products
to be sterilized. |
| Reference
standard: |
Standard,
generally having the highest metrological quality available at a given
location or in a given organisation, from which measurements made there
are derived. |
| Safety
hazard: |
Potentially
detrimental effect on persons or the surroundings arising directly from
either the sterilizer or its load. |
| Saturated
steam: |
Water
vapour at a temperature corresponding to the boiling point of the source
liquid. |
| Small
steam sterilizers: |
Steam
sterilizer which is unable to accommodate a sterilization module. |
| Sterile: |
Condition
of a medical device that is free from viable micro-organisms. |
| Sterilization: |
Process
undertaken to render a sterilizer load sterile. |
| Sterilization
cycle: |
Automatic
sequence of operating stages performed in a sterilizer for the purpose
of sterilization. |
| Sterilization:
temperature: |
Minimum
temperature of the sterilization temperature band. |
| Sterilization
temperature band: |
Range
of temperatures, expressed as the sterilization temperature and the
maximum allowable temperature which may prevail throughout the load
during the holding time.
NOTE:
These temperatures are usually stated in whole degrees Celsius. |
| Sterilizer: |
Apparatus
designed to achieve sterilization. |
| Sterilizer
chamber: |
That part of the sterilizer which receives the sterilizer load. |
| Sterilizer
load: |
Goods
that are to be sterilized simultaneously in the same sterilizer chamber. |
| Superheated
steam: |
Steam
whose temperature, at any given pressure, is higher than that indicated
by the vaporisation curve of water. |
| Test
organism: |
Micro-organisms
used for the manufacture of inoculated carriers. |
| Type
test: |
Series
of tests to establish the working data for a sterilizer type. |
| Unloading
door: |
Door
in a double ended sterilizer through which the sterilized load is removed
from the sterilizer chamber after a sterilization cycle. |
| Usable
space: |
Space
inside the sterilizer chamber which Is not restricted by fixed parts
and which is consequently available to accept the sterilizer load. NOTE: The usable space is expressed in terms of chamber height, chamber width and chamber depth. |
| Validation: |
Documented
procedure for obtaining, recording and interpreting data required to
show that a process will consistently comply with predetermined specifications. |
| Works
test: |
Series
of tests performed at the manufacturer's works to demonstrate compliance
of each sterilizer with its specification. |